전리함 반응 함수의 직접 측정과 이를 이용한 방사선의 실제선량 분포측정

Direct Measurement of Chamber Response Function and Its Application to Radiation Dose Distribution Dosimetry

  • 13
커버이미지 없음

Purpose : To obtain the actual dose distribution from measured data by deconvolution method using the measured ion chamber response function. Materials and Methods : The chamber response functions for 2 ionization chambers (diameter 5mm, 6.4mm) were measured, and dose profiles were measured for 10×20cm2 field size using two different detectors. The deconvolution of chamber response function from the measured data were performed for these profiles. The same procedures were repeated for 4MV, 6MV and 15MV photon energies. Results : Different dose profiles were obtained for the same field with the chambers which have the different response functions. Nearly the same results could be obtained with deconvolution for the profiles from various detectors. Conclusion : The effect of the chamber response function can be extracted by deconvolution method. Deconvolved dose profile using various ionization chambers gave better dose distributions. Technical improvements are needed for practical application.

목 적 : 전리함의 크기로 인한 공간 분해능의 문제로 나타나는 전리함 반응함수를 제거하여 실제 선량분포를 얻고자 하였다. 대상 및 방법 : 내경 5mm, 6.4mm 등 2개의 서로 다른 크기를 갖는 전리함들과 다이오드, 필름의 반응함수를 구하고, 동일한 방사선 조사면(10×20cm2)의 선량분포 프로파일을 측정하여, 각각 deconvolution 기법으로 보정한 후, 보정된 결과가 일치하는지 비교하였다. 결 과 : 원통형 전리함의 반응함수와 선량분포를 측정하였고, 측정한 선량분포에서 반응 함수의 효과를 deconvolution방법으로 제거하여 실제 선량분포를 찾아내었다. 사용한 에너지는 최대 광자선 에너지 4MV, 6MV, 15MV였으며, 전 에너지 영역에 걸쳐 보정 된 결과가 일치하는 방향으로 변화하였으며, 분해능 증가 효과가 있었다. 결 론 : deconvolution 방법으로 전리함 반응 함수의 효과를 제거했을 때, 여러 측정기를 이용하여 측정한 선량 프로파일의 결과가 일치하는 경향을 볼 수 있어서 deconvolution 방법을 통해 얻은 선량 프로파일을 임상적으로 응용할 수 있음을 알았다

(0)

(0)

로딩중