부정형 조사면에서의 TLD를 이용한 방사선 흡수선량 측정

Dosimetry of Irregular field Using Thermoluminescence Dosimetry

  • 11
커버이미지 없음

In clinical radiotherapy, the use of wide and irregular field techniques frequently results in considerable tumor dose inhomogeneity because of, the variation in physical characteristics of irradiated volumes. This report describes an analysis of the dosimetry of the irregular fields such as radaition fields for Hodgkin's disease(mantle field), esophageal cancer, and lung cancer when a 6MV and a 15MV linear accelerators are utilized. Doses were measured in a Rando phantom using methods of thermoluminescence dosimetry(TLD) and were calculated by radiotherapy planning computer system with the Clarkson's method for calculation of a irregular field. A dose variation of 5-22%, 6-9%, 6-14% were found in the mantle field, esophageal cancer field, lung cancer field respectively. Higher doses occurred in the superior portion of the irregular field. The sites of maximum dose variation were the supraclavicular and the upper spinal cord region. To adjust for these substantial differences, a compensator or a shrinking field technique should be adopted.