Accuracy of Dose Estimation in High Dose Rate Intracavitary Radiotherapy of Carcinoma of the Uterine Cervix
- 대한방사선종양학회
- 대한방사선종양학회지
- 제5권 제2호
-
1987.12137 - 139 (3 pages)
- 10
커버이미지 없음
In brachytherapy of uterine conical cancer using a high dose rate remote afterloading system, it is of prime importance to deliver a accurate dose in each fractionated treatment by minimizing the difference between the pre-treatment planned and post-treatment calculated doses. The post-treatment calculated point A dose was not much different from the pretreatment planned dose (500 cGy). The average±standard deviation was 500±18cGy and 84 percent of 82 intracavitary radiotherapy was within the range of 500±25cGy.