고효율의 ESI-TOF/Ms 시스템을 이용한 생리활성 항체와

방사성동위원소 표지용 착화제의 결합 검증

  • 7
114977.jpg

In this study, we measured the complex efficiency of a physiologically active antibody, a chelator and radi- osiotopes using the ESI-TOF/Ms system for develop good radiopharmaceuticals. For a precise measurement, TLC is a low accuracy method. Loading of same amount of sample is difficult for each test, and work to quantify accurately the results obtained through TLC cannot be afforded compared to the use of other analytical instruments. The method of analysis using a mass spectrometer is capable of a mass analysis of proteins for quantitative analysis. The conjugates of the chelator (CHX- A-DTPA) and the antibody (IgG) were separated through MWCO, and were analyzed using ESI-TOF and MALDI-TOF mass spectrometry. The analysis using MALDI-TOF is roughly divided into measurements on mass spectrometry. When con- jugating a small molecular weight of CHX-A-DTPA and a large molecular weight of IgG, distinguishing the peak of the con- jugate and the peak of IgG was difficult. However, an ESI-TOF mass spectrometer system is capable of an analysis of mass by decentralizing the IgG. It is utilized as a technique for measuring the metabolic processes during conjugation and the stability evaluation of radiopharmaceuticals. When establishing this technique, the accuracy of the overall radiophar-maceutical analysis is expected to be able to be improved.