H+ 선택성 막 전극을 이용한 바이오센서의 감응 특성

Characteristic Responses of Biosensor Based on H+-Selective Membrane Electrode

  • 20
116387.jpg

The H+-selective membrane consists of tridodecylamine, 2-nitrophenyl octyl ether and a little amount of additive in carboxylated PVC matrix, where penicillinase(Pcase) and glutaraldehyde may be covalently attached to the matrix. When the H+-selective electrode was used as a detector of biosensor, calibration curve calculated from Nernst equation was not linear. So we investigated the characteristic responses of the H+-selective electrode to the product H+ from hydrolysis of penicillin-G(Pc-G) and plotted calibration curve correlating potential to concentration of Pc-G linearly. The optimal concentration of buffer solution was theoretically calculated and was also experimented. We tried to explain the linear curve of potential to concentration of Pc-G by using Henderson-Hasselbach equation. This method is more effective in calibration curve plotting than any other previous methods. The results obtained may help in further developing pH electrodes with improved analytical preformance.