상세검색
최근 검색어 전체 삭제
다국어입력
즐겨찾기0
학술저널

녹내장 의증 환자의 감별에서 가변각막보정 주사레이저편광측정의 진단적 가치

GDx-VCC Performance to Discriminate Normal, Pre-perimetric Glaucomatous Eyes

  • 2
121424.jpg

Purpose: This study was designed to assess the diagnostic value of scanning laser polarimetry with variable corneal compensation (GDx-VCC) in the diagnosis of preperimetric glaucoma. Methods: The study included 132 eyes of 132 patients with normal intraocular pressure, including 38 normal eyes, 60 eyes with pre-perimetric glaucoma, and 34 eyes with early glaucoma. The parameters of GDx-VCC were analyzed and compared in these groups using ANOVA. The parameter with the most powerful diagnostic value was defermirred by an ROC curve, and it`s sensitivity and specificity were calculated. Results: Among GDx-VCC parameters, NFI was the most valuable parameter that could detect pre-perimetric glaucoma in normal eyes. A cut-off value of 12 was the optimal NFI value, which offered the highest sensitivity and specificity in discriminating between normal and pre-perimetric glaucomatous eyes. However, there were no statistically significant differences in GDx-VCC parameters between pre-perimetric and early glaucomatous eyes. A statistically significant correlation was found between NFI and visual field indices in early glaucomatous eyes. Conclusions: GDx-VCC can be useful as a screening test for early detection of pre-perimetric and early glaucoma with normal intraocular pressure.

(0)

(0)

로딩중