서로 다른 비구면 인공수정체 삽입안에서 술 후 시기능 비교
Comparison of Visual Function Among Aspheric Intraocular Lenses
- 대한안과학회
- 대한안과학회지
- Ophthalmological Society,volume50,number5
-
2009.05691 - 697 (7 pages)
- 0
Purpose: To compare the visual function among 3 aspheric intraocular lenses (IOLs) after cataract surgery. Methods: A prospective analysis among 70 eyes of 68 patients who had undergone phacoemulsification and implantation of aspheric IOLs (Acrysof IQ, 30 eyes; Tecnis ZA9003, 20 eyes; Akreos Adapt-AO, 20 eyes) was performed. The best corrected visual acuity (BCVA), refraction, coma, trefoil, spherical aberration, root-mean-square (RMS), contrast sensitivity test, satisfaction and glare was evaluated during the follow-up period of 3 months. Results: There were no significant differences in BCVA, satisfaction, glare, coma, trefoil, RMS and contrast sensitivity test among the 3 aspheric IOLs. Spherical aberration in Akreos Adapt-AO (0.04±0.05 μm) was higher compared with Acrysof IQ (0.03±0.04 μm) and Tecnis ZA9003 (0.11±0.03 μm)(p<0.01). The predictability of postoperative spherical equivalent in the Akreos Adapt-AO group (-0.57±0.22D) represented a statistically significant myopic refractive error compared with the Acrysof IQ (0.01±0.77D) and the Tecnis ZA9003 (-0.20±0.71D) groups (p=0.04). Conclusions: There were no significant differences in visual function except spherical aberration, which was highest in the Akreos Adapt-AO group.