하이드로겔 재질의 전(全)광학부 형태인 조절인공수정체의 임상결과와 안정성

Visual Outcome and Stability of Hydrogel Full-Optics Accommodative Intraocular Lens

  • 9
121474.jpg

Purpose: To report the visual outcome and to determine the stability of WIOL-CF? (GELMED, Praha, Czech Republic) hydrogel full-optics accommodative intraocular lens (IOL). Methods: The present study evaluated 20 eyes of 10 patients who underwent routine cataract surgery and WIOL-CF? accommodative IOL implantation. Measurement included uncorrected/best corrected visual acuities (VA) at near and distant, degree of IOL decentration, degree of IOL tilting and anterior chamber depth on postoperative day 1, 1 month, 2 months, 6 months and 12 months. Results: Result analysis was performed with 19 eyes of 10 patients, except for 1 complicated eye. At 12 months, the uncorrected distance VA of 11 eyes (57.9%) and 17 eyes (89.5%) were 20/25 and 20/40 or better, respectively. At 12 months, the uncorrected near VA of 12 eyes (63.2%) and 18 eyes (94.7%) were 20/25 and 20/40 or better, respectively. There was no statistically significant difference in anterior chamber depth (p > 0.05), IOL decentration (p > 0.05) and IOL tilting (p > 0.05) on postoperative day 1, 1 month, 2 months, 6 months and 12 months. Conclusions: The clinical results of cataract surgery with WIOL-CF were relatively satisfactory. J Korean Ophthalmol Soc 2011;52(12):1448-1454