이중샤임플러그, 전안부 빛간섭단층촬영계 및 초음파 각막두께측정계를 이용한 각막두께의 비교

Corneal Thickness Measured by Dual Scheimpflug, Anterior Segment Optical Coherence Tomography, and Ultrasound Pachymetry

  • 0
121484.jpg

Purpose: To compare central corneal thickness (CCT) as measured by dual rotating Scheimpflug camera (Galilei), anterior segment optical coherence tomography (AS-OCT), and ultrasound pachymetry (USP). Methods: The measurements of CCT using a dual rotating Scheimpflug camera, AS-OCT, and USP in 40 eyes of 20 healthy subjects were compared. Results: The average measurements of CCT by dual rotating Scheimpflug camera, AS-OCT, and USP were 538.10 ± 31.36 μm, 536.20 ± 31.21 μm, and 541.93 ± 34.93 μm, respectively. The CCT measurement by USP was statistically significantly thicker than by the dual rotating Scheimpflug camera and AS-OCT (p = 0.017, p = 0.001, respectively). There was no significant difference between the dual rotating Scheimpflug camera and AS-OCT (p = 0.054). A significant linear correlation was observed between the dual rotating Scheimpflug camera, the AS-OCT, and the USP (r > 0.900, p < 0.001). Conclusions: The results of the 3 methods have significant correlation with each other, but the measurement by USP was significantly thicker than the dual rotating Scheimpflug camera and AS-OCT. Therefore, CCT should be interpreted in the context of the instrument used. J Korean Ophthalmol Soc 2012;53(10):1412-1418