각막굴절교정수술 후 펜타캠을 이용한 안압 보정방법의 비교
Comparison of Intraocular Pressure Correction Programs in Pentacam after Corneal Refractive Surgery
- 대한안과학회
- 대한안과학회지
- Ophthalmological Society,volume54,number1
-
2013.0126 - 32 (7 pages)
- 13
Purpose: To evaluate the accuracy of Pentacam? built-in 5 intraocular pressure (IOP) correction programs used to measure the IOP of patients who received corneal refractive surgery. Methods: IOP of 124 eyes from 62 patients who underwent epipolis laser in situ keratomileusis was measured with Goldmann applanation tonometry (GAT) at 6 months pre- and post-operatively. The collected data was input into Pentacam?, calculated by 5 correction programs, Ehlers, Shah, Dresden, Orssengo / Pye, Kohlhaas, and compared. Results: The GAT-based pre- and post-operative IOP was 15.75 ± 2.24 mm Hg, and 10.72 ± 2.31 mm Hg, respectively, revealing the post-operative IOP to be significantly lower than the pre-operative IOP (p < 0.001). Among the 5 correction programs within Pentacam?, Ehlers program showed little difference between pre- and post-operative IOP values (p = 0.228) and the post-operative correction value showed no significant difference with the pre-operative GAT value (p = 0.413). Conclusions: The Ehlers program is the most accurate among the 5 Pentacam? correction programs evaluated in the present study, and can be a useful tool for correcting the true IOP of patients which tends to be higher after corneal refractive surgery.
(0)
(0)