접촉식 초음파와 부분결합간섭계를 이용한 생체계측시 백내장수술 후 굴절력 예측의 비교

Comparison of the Refractive Results Measured by Ultrasound and Partial Coherence Interferometers

  • 0
121502.jpg

Purpose: To compare ocular biometry measured by applanation ultrasonography and IOL Master<sup>&reg;, and evaluate the accuracies of the refractive outcome after cataract surgery. Methods: The biometries of 76 cataractous eyes were measured using ultrasonography and IOL Master<sup>&reg;. The SRK-T formula was employed to predict the patient’s implanted IOL power. Two months after cataract surgery, the refractive outcome was determined, and results from the 2 different biometry methods were compared. Results: There were no statistically significant differences in axial length (AXL) and anterior chamber depth (ACD) between ultrasonography and IOL Master<sup>&reg; (p = 0.501). When using ultrasonography, the mean absolute error (MAE) of the ultrasonography and IOL Master<sup>&reg; was 0.53 ± 0.30 diopter (D), 0.55 ± 0.41 D, respectively. The difference between the 2 biometry methods was not statistically significant (p = 0.110). Conclusions: Although the difference was not statistically significant, AXL measured by IOL Master<sup>&reg; was longer and ACD measured by IOL Master<sup>&reg; was deeper than when measured by A-scan. This difference was more pronounced in patients with a short AXL. The accuracy of IOL power calculation was similar between the 2 devices.