Formula Comparison for Intraocular Lens Power Calculation Using IOL Master and Ultrasound for the ZCB00 IOL
- 대한안과학회
- 대한안과학회지
- Ophthalmological Society,volume55,number4
-
2014.04527 - 533 (7 pages)
- 2
Purpose: To evaluation the accuracy of the IOL power calculation formulae measured by IOL Master<sup>® and applanation ultrasonography for the Tecnis ZCB00 IOL Methods: We performed a retrospective study of 170 eyes in 121 patients who underwent cataract surgery in our hospital with AMO Tecnis ZCB00 IOL.s. The SRK/T formula was used to predict the patient's implanted IOL power. Differences in the predicted refractive errors between IOL Master<sup>® and ultrasonography were analyzed and factors attributed to the differences were also analyzed. Three months after cataract surgery, mean numeric error and mean absolute error were analyzed. Results: SRK/II and SRK/T formulas calculated using ultrasonography showed differences compared to the same formulas calculated using IOL Master<sup>®, in which hyperopic shift was also demonstrated. No definite factor was attributed to the differences between the 2 methods. Although the 3 formulas of IOL Master<sup>® showed no significant difference in refractive errors, the SRK/T formula calculated using IOL Master<sup>® showed the least mean absolute and numeric errors. Conclusions: IOL Master<sup>® is considered more suitable when determining proper AMO Tecnis ZCB00 IOL power in cataract surgery. The hyperopic shift should be considered when calculating the IOL power using only ultrasonography. J Korean Ophthalmol Soc 2014;55(4):527-533