수평각막직경을 포함한 전안부 계측에 있어서 슬릿광주사 각막형태검사기와 저간섭성 반사계의 비교

Comparison of Anterior Segment Measurements Using Scanning-Slit Topography and Optical Low-Coherence Reflectometry (OLCR) Biometry

  • 5
121514.jpg

Purpose: To compare the results of anterior segment biometry including white-to-white (WTW) between scanning-slit topography (ORBscan IIz<sup>&reg;, Bausch & Lomb), optical low-coherence reflectometry (OLCR) biometry (Lenstar<sup>&reg;, Haag-Streit), and Castroviejo calipers. Methods: Measurements on 72 eyes of 36 patients that underwent refractive surgery were measured using ORBscan<sup>&reg;, Lenstar<sup>&reg;, and calipers and compared. Ocular biometry parameters used in this study included the WTW, central corneal thickness, anterior chamber depth (ACD), keratometry, and pupil size. Results: The WTW measurements using ORBscan<sup>&reg; and calipers (11.57 ± 0.35 mm and 11.58 ± 0.34 mm, respectively) were statistically similar. However, the measurement using Lenstar<sup>&reg; (12.05 ± 0.40 mm) was significantly greater than with the other methods (<em>p</em> < 0.001). Central corneal thickness and keratometry measurements using ORBscan<sup>&reg; were greater than when using Lenstar<sup>&reg; (<em>p</em> = 0.01 for both). ACD and pupil size measurement using Lenstar<sup>&reg; were greater than when using ORBscan<sup>&reg; (<em>p</em> < 0.001 for both). Conclusions: Because WTW and ACD measurements using Lenstar<sup>&reg; were greater than when using ORBscan<sup>&reg; and calipers, unexpected high-vaulting may be observed due to the selection of a larger-sized posterior chamber phakic intraocular lens. Therefore, the differences in measurements obtained when using these methods should be considered. J Korean Ophthalmol Soc 2014;55(5):656-661

(0)

(0)

로딩중