Objectives: This study was performed to evaluate the analytical method for PAH metabolites in human urine using enzyme hydrolysis and solid-phase extraction coupled with LC-(ESI)-MS/MS technique. Methods: We employed HPLC tandem mass spectrometry techniques with appropriate pre-treatment for analysis of 16 OH-PAHs in human urine. Samples were hydrolysis by β-flucuronidase/Aryl sulfatase, and target compounds were extracted by solid-phase extraction with a strata-x cartridge. Cross-validation was performed between Eulji University and Green Cross laboratories with 200 human urine samples. Results: The accuracies were between 90.3% and 118.8%, and precisions (relative standard deviations) were lower than 10%. The linearity obtained was satisfying for the 16 OH-PAH compounds, with a coefficient of determination (r 2 ) higher than 0.99. The results of cross-validation at the two organizations were compared by ICC (interclass correlation coefficient) values. The cross-validation results were excellent or good for all compounds. Conclusion: An analytical method was validated for low nanogram levels of 16 OH-PAHs in human urine. Also, satisfying results were obtained for method validation such as accuracy, precision and ICC of cross-validation.
I. 서 론
II. 재료 및 방법
III. 연구결과
IV. 고 찰
V. 결 론
(0)
(0)