Biological Assay of Mercury and Cadmium Ions Using DNA Immobilized on a Nanotube Paste Electrodes
- 한국응용과학기술학회 (구.한국유화학회)
- 한국응용과학기술학회지
- 한국유화학회지 제29권 제2호
-
2012.06302 - 310 (9 pages)
- 13
Bio assay of mercury and cadmium ions were searched using voltammetric analysis using DNA doped carbon nanotube paste electrodes (DCP). The square-wave stripping voltammetryic optimized results indicated working ranges of 1-10.0 ngL-1 and 20-100ugL-1,Hg(II) Cd(II) within an accumulation time of 120 seconds, in 0.1-M phosphate buffer solutions of pH 6.3. The relative standard deviations of 5 ngL-1 Hg(II) and Cd(II) that observed were 0.14 and 0.22% (n=12), respectively, using optimum conditions. The low detection limit(S/N) was pegged at 0.1 ngL-1 (4.9×10-11M) Hg(II) and 0.2 ngL-1 (1.77×10-10M) Cd(II). The developed methods can be applied to assays in biological fish kidneys and water samples
1. Introduction
2. Experimental Method
3. Results and Discussion
4. Conclusions
References