호흡음주측정에서 Alco-Sensor Ⅳ와 Lion Alcolmeter SD-400P의 비교
The Comparison of Alco-Sensor Ⅳ and Lion Alcolmeter SD-400P in Breath Alcohol Measurements
- 한국중독정신의학회
- 중독정신의학
- Vol.3, No.1
-
1999.0468 - 74 (7 pages)
- 11
Objectives:The aim of this study was to evaluate the interreliabilities and reproducibilities of two breathanalyzers which commonly used in clinical settings, research fields, and legal purposes including roadside testing. Methods:Thirty-four healthy male volunteers were examined through 30, 60, 90, 120, 150 and 180 minutes after 20 minutes’ intake of one of the two doses of alcohol (0.5 and 0.75 g/kg of body weight) on separate occasions. Two kinds of breathanalyzers [Alco- Sensor Ⅳ (Intoximeters, Inc., USA;Alco-Sensor) and Lion Alcolmeter SD-400P (Lion Laboratories plc, UK;SD-400P)] were used to measure the breath alcohol concentrations (BrAC). Duplicate breath samples at each time point were used to determine the repeatabilities of the breathanalyzers. Results:Mean values of two breathanalyzers at each time point were not statistically different. Although there were not significant, the values obtained from SD-400P were tended to slightly higher than those of Alco-Sensor. Inter-reliabilities of Alco-Sensor and SD-400P at 0.5 and 0.75 g/kg of body weight were very high[r=0.972 (p<.001);r= 0.938 (p<.001), respectively]. Test-retest reliabilities of Alco-Sensor and SD- 400P were also very high[r=0.976 (p<.001); r=0.984 (p<.001), respectively]. Conclusion:The two breathanalyzers can be used as highly reliable instruments in all kinds of fields including roadside testing. Although further studies concerning the accuracy and precision of the breath-alcohol measurements in Koreans should be done, it suggests that these instruments may be easily and safely used as evidential breath testing.
서 론
연구 대상 및 방법
결 과
고 찰
결 론
(0)
(0)