상세검색
최근 검색어 전체 삭제
다국어입력
즐겨찾기0
학술저널

소프트웨어 기반 정도관리 시스템을 이용한 부피세기조절회전치료 환자 별 정도관리의 유용성 평가

Evaluation of the Utility of a Volumetric Modulated Arc Therapy Specific Patient Quality Assurance using Software-based Quality Assurance System

  • 37
138346.jpg

The purpose of this study is to evaluate the usefulness of a software-based quality assurance system based on Volumetric Modulated Arc Therapy treatment plan. Evaluate treatment plan through the D VH analysis, PTV mean dose (   ) and PTV 95% dose(   ) compare the MFX based on original treatment plan, Average error rate was 0.9±0.6%, 1.0±0.8%, respectively. Measuring point dose using phantom and ion chamber, the average error rate between the ionization chamber and MFX was 0.9±0.7%, 1.1±0.7% (high dose region), 1.1±0.9%, 1.2±0.7% (low dose region). The average gamma though of MFX and Delta 4PT is 98.7±1.2%, 98.4±1.3%, respectively. Through this study, A software based QA system that simplifies hardware based QA procedures that involve a lot of time and effort. It can be used as a simple and useful tool in clinical practice.

Ⅰ. 서 론

Ⅱ. 대상 및 방법

Ⅲ. 결 과

Ⅳ. 고찰 및 결론

(0)

(0)

로딩중