소프트웨어 기반 정도관리 시스템을 이용한 부피세기조절회전치료 환자 별 정도관리의 유용성 평가
Evaluation of the Utility of a Volumetric Modulated Arc Therapy Specific Patient Quality Assurance using Software-based Quality Assurance System
- 대한방사선과학회(구 대한방사선기술학회)
- 방사선기술과학
- 방사선기술과학 제41권 제1호
-
2018.021 - 8 (8 pages)
- 37
The purpose of this study is to evaluate the usefulness of a software-based quality assurance system based on Volumetric Modulated Arc Therapy treatment plan. Evaluate treatment plan through the D VH analysis, PTV mean dose (   ) and PTV 95% dose(   ) compare the MFX based on original treatment plan, Average error rate was 0.9±0.6%, 1.0±0.8%, respectively. Measuring point dose using phantom and ion chamber, the average error rate between the ionization chamber and MFX was 0.9±0.7%, 1.1±0.7% (high dose region), 1.1±0.9%, 1.2±0.7% (low dose region). The average gamma though of MFX and Delta 4PT is 98.7±1.2%, 98.4±1.3%, respectively. Through this study, A software based QA system that simplifies hardware based QA procedures that involve a lot of time and effort. It can be used as a simple and useful tool in clinical practice.
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고찰 및 결론
(0)
(0)