저유량 마취를 위한 전자제어식 기화기 개발
A Development of the electronic controlled vaporizer for low flow anesthesia
- 강원대학교 산업기술연구소
- 산업기술연구
- vol.29
-
2009.09107 - 114 (8 pages)
- 10
For the environmental safety of the operating room and patient healthcare, the closed type rebreathing system is widely adopted. In order to reduce the anesthesia gas during surgery, the mixing ratio of anesthesia gas with breathing air should be precisely controlled. Generally, the breathing air passes through the vaporizer to mix the anesthesia gas, but there is a difficulty in controlling the mixing ratio precisely. In this paper, the stand-alone style vaporizer is designed and the operating characteristics are investigated. The vaporizer measures the temperature and pressure in the vaporizing chamber and chamber temperature is precisely controlled by proportional controlled heater. Exact quantity of anesthesia media is feeded by PID controlled peristaltic pump and vaporized gas is mixed with breathing air flow by PWM controlled solenoid valve. The experimental result shows that the vaporizer has an excellent command following performances that it can be applied to the low flow anesthesia system.
1. 서론
2. 이론적 배경
3. 연구 방법
4 성능실험
5. 결론
6. 후기
(0)
(0)