늑골 방사선검사 시 X선관 초점-영상수신체간 거리에 따른 환자선량과 화질의 연관성 평가
Evaluation of Relationship between Radiation Dose and Image Quality according to Source to Image Receptor Distance in Rib Series Radiography
- 주영철(Young-Cheol Joo) 정영진(Young-Jin Jung)
- 대한방사선과학회(구 대한방사선기술학회)
- 방사선기술과학
- 방사선기술과학 제41권 제5호
- 등재여부 : KCI등재
- 2018.10
- 391 - 396 (6 pages)
The purpose of this study was to compare and analyze the patient dose according to the distance between the X-ray tube focus and the image receptor, and to propose a new method for quantitatively evaluating the image quality. Using this quantitative evaluation method, the optimal distance for increasing x-ray image quality with low radiation dose was estimated between source and image receptor in Rib series radiography. Phantom images were obtained by changing the distance between focus and image receptor (100 ㎝ and 180 ㎝). The patient radiation dose was estimated using en-trance surface dose and dose area product. In order to evaluate image quality objectively, a non - reference image evaluation method was employed with paper and salt noise and Gaussian filter. As a result of this study, when the SID was changed from 100 ㎝ to 180 ㎝, the entrance surface dose decreased by 4 ~ 5 times and the dose area product decreased by 3 times. In addition, there is no significant difference in image quality between of SID 180 cm and SID 100 cm. In conclusion, it was demonstrated that performing the rib series radiography at SID 180 cm is an optimal method to reduce the exposure dose and improve the image quality.
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론