수도권 지역의 초음파 프로브의 미생물 오염도와 소독에 관한 연구
A Study on Microbial Contamination and Disinfection of Ultrasonic Probe in Metropolitan Area
- 대한방사선과학회(구 대한방사선기술학회)
- 방사선기술과학
- 방사선기술과학 제41권 제5호
-
2018.10427 - 435 (9 pages)
- 27
There was a shortage of research reports on sterilization criterion and contamination of ultrasonic probes. Therefore, in this study, we were going to provide a basic study to measure the level of microbial contamination in ultrasonic probes and to investigate the radiographer s awareness of infection. After the scan, samples were collected from the rubber part of the probe by opening a sterile swab (Transport Medium AM608-1S) for medical bacteria collection with the remaining gel removed with a paper towel. Also, the collected samples of bacteria were grown for seven days and then the laboratory was analyzed. Among the total 29 types of microorganisms, Micrococcus luteus 21(26%), Moraxella species 16(20%), Coagulase negative staphylococcus 8(10%), Bacillus species 5(7%), Bicillus circulans 3(5%), Acinetobacter lwoffii 2(2%), and 1 other Candida parapsilosis (1%) a number of bacteria and fungus, was detected. In a disinfectant experiment using LuciPac Pen on the Lumitester PD-30s, we cultured the rubber part of the probe two to three times to measure the bacteria. Bacteria decreased to 97% with Aquanax (alkaline reduced water 100%), 99% with Klarion wash (0.01% sodium hydroxide), 94% with Klarion disinfection (0.01% nitrous acid water), Sterilization was best with Klarion wash (0.01% sodium hydroxide). Therefore, guidelines for cleaning and disinfection of ultrasonic probes was required, and further development of probe-only disinfectants is required.
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
(0)
(0)