Diagnostic ex vivo assay of glucose in live cell using voltammetry
- 한국응용과학기술학회 (구.한국유화학회)
- 한국응용과학기술학회지
- 한국유화학회지 제35권 제4호
-
2018.121379 - 1385 (7 pages)
- 11

The hand held voltammetry systems searched diabetic assay using glucose sensor of fluorine nafion doped carbon nanotube electrode (FCNE). An inexpensive graphite carbon pencil was used as an Ag/AgCl reference and Pt counter electrode. Upon combining and using three electrode systems, optimum square wave (SW) stripping results were attained to 1.0-9.0 ug/L with 8 points. Statistic RSD precision was of 6.02 % with n=15 in 0.1 mg/L glucose. After a total of 200 second accumulation times, analytical detection limit of 0.8 ug/L was obtained. This developed technique was applied to urine samples from diabetic patients urine for fluid analysis, it was determined that the sensor can be used with a diagnostics in the ex vivo of live cells and non treated biological fluid.
Abstract
1. Introduction
2. Experimental procedure
3. Results and discussion
4. Conclusion
References