Effect of Cataract Grade according to Wide-Field Fundus Images on Measurement of Macular Thickness in Cataract Patients
- Mingue Kim Youngsub Eom Jong Suk Song Hyo Myung Kim
- 대한안과학회
- The Korean Journal of Ophthalmology
- Vol.32 No.3
- 등재여부 : KCI등재
- 2018.06
- 172 - 181 (10 pages)
Purpose: To investigate the effects of cataract grade based on wide-field fundus imaging on macular thickness measured by spectral domain optical coherence tomography (SD-OCT) and its signal-to-noise ratio (SNR). Methods: Two hundred cataract patients (200 eyes) with preoperative measurements by wide-field fundus imaging and macular SD-OCT were enrolled. Cataract severity was graded from 1 to 4 according to the degree of macular obscuring by cataract artifact in fundus photo images. Cataract grade based on wide-field fundus image, the Lens Opacity Classification System III, macular thickness, and SD-OCT SNR were compared. All SD-OCT B-scan images were evaluated to detect errors in retinal layer segmentation. Results: Cataract grade based on wide-field fundus imaging was positively correlated with grade of posterior subcapsular cataracts (rho = 0.486, p < 0.001), but not with nuclear opalescence or cortical cataract using the Lens Opacity Classification System III. Cataract grade was negatively correlated with total macular thickness (rho = -0.509, p < 0.001) and SD-OCT SNR (rho = -0.568, p < 0.001). SD-OCT SNR was positively correlated with total macular thickness (rho = 0.571, p < 0.001). Of 200 eyes, 97 (48.5%) had segmentation errors on SD-OCT. As cataract grade increased and SD-OCT SNR decreased, the percentage of eyes with segmentation errors on SD-OCT increased. All measurements of macular thickness in eyes without segmentation errors were significantly greater than those of eyes with segmentation errors. Conclusions: Posterior subcapsular cataracts had profound effects on cataract grade based on wide-field fundus imaging. As cataract grade based on wide-field fundus image increased, macular thickness tended to be underestimated due to segmentation errors in SD-OCT images. Segmentation errors in SD-OCT should be considered when evaluating macular thickness in eyes with cataracts.
Materials and Methods
Results
Discussion
Conflict of Interest
Acknowledgements
References