상세검색
최근 검색어 전체 삭제
다국어입력
즐겨찾기0
학술저널

Non-contact Tonometry: an Ideal Method for Mass Screening

  • 0
146355.jpg

We evaluated one of the new non-contact tonometers, the Pulsair non-contact tonometer, to assess its accuracy and reliability. Measurements on 101 eyes were used to assess its accuracy against Goldmann tonometry measurements, and another 24 eyes were used to assess its reliability (reproducibility). The results showed the instrument to be highly accurate and reliable in measurement.

Abstract

INTRODUCTION

MATERIALS AND METHODS

RESULTS

DISCUSSION

REFERENCES

(0)

(0)

로딩중