Whole brain radiotherapy using four-field box technique with tilting baseplate for parotid gland sparing
- 대한방사선종양학회
- 대한방사선종양학회지
- 제37권 제1호
-
2019.0322 - 29 (8 pages)
- 0

Purpose: The aim of this study is to evaluate the efficacy and feasibility of four-field box whole brain radiotherapy (FB-WBRT) with tilting baseplate by comparing bilateral WBRT (B-WBRT). Methods and Materials: Between March 2016 and September 2018, 20 patients with brain metastases underwent WBRT using the four-field box technique. WBRT is performed with a dose of 30 Gy in 10 fractions daily. Two computed tomography simulations per person were performed. One was in the traditional supine position for B-WBRT and the other by applying the tilting acrylic supine baseplate to elevate the head by 40° for FB-WBRT. The B-WBRT used the field-in-field technique, which is the most commonly used method in our institution. The FB-WBRT comprised anterior, posterior, and bilateral beams. A wedge was applied in anterior and posterior fields to compensate for skull convexity. Results: The average of Dmean of both parotid glands was 10.2 Gy (range, 3.8 to 17.8 Gy) in B-WBRT and 5.4 Gy (range, 2.0 to 11.7 Gy) in FB-WBRT (p < 0.05). Compared to B-WBRT, FB-WBRT reduced the mean dose of the right and left parotid glands from 10.1 Gy to 4.9 Gy and from 10.4 Gy to 5.8 Gy, respectively (p < 0.05). Further, V5, V10, V15, V20, and V25 for the parotid gland decreased significantly in FB-WBRT (p < 0.05). The Dmax and Dmean of lens decreased according to the dose-volume histogram. Conclusion: Compared to B-WBRT, FB-WBRT with a tilting baseplate is a simple and effective method that takes feature of noncoplanar beam to protect the parotid gland.
Introduction
Materials and Methods
Results
Discussion and Conclusion
Conflict of Interest
References
(0)
(0)