Reduction of Radiation Dose for Injection of Radioisotope using Shielding Device

Reduction of Radiation Dose for Injection of Radioisotope using Shielding Device

  • 19
※해당 콘텐츠는 기관과의 협약에 따라 현재 이용하실 수 없습니다.

Nuclear medicine have often used to diagnose cancers. The main absorbed dose from radiation to a radiation worker resulted from open radioisotopes. Methods for reducing the radiation dose to a radiation worker from radioisotopes injected to patients were studied. The shield device of 0.2 mmPb was manufactured as a size of 300 mm × 500 mm × 150 mm. By using dosimeters of Nanodot, the absorbed doses for thyroid, chest and genital organ were measured with and without a shielding device and with syringe shield and shielding device together. The highest absorbed dose of 0.908 mGy reduction of 20.8% as 0.719 mGy was in the genital organ by using the syringe shield and a shielding device together. A effective dose for a radiation worker during 1 year was expected to 1.223 mSv at the chest, which was decrease as 0.994 mSv by shielding device and syringe shield together. When open radioisotope is injected to a patient for examination, the only use of a shielding device results in the reduction of radiation dose to radiation workers.

Ⅰ . INTRODUCTION

Ⅱ. MATERIAL AND METHODS

Ⅲ. RESULT

Ⅳ. DISCUSSION

Ⅴ. CONCLUSION