인터벤션 방사선발생장치에서 입사표면선량률 평가
Evaluation of Radiation Entrance Surface Dose Rates for Interventional Radiology Equipment
- 대한방사선과학회(구 대한방사선기술학회)
- 방사선기술과학
- 방사선기술과학 제43권 제5호
-
2020.10353 - 357 (5 pages)
-
DOI : 10.17946/JRST.2020.43.5.353
- 12
IVR procedures are on the rise, and patient doses are on the rise. It is necessary to evaluate fluoroscopy dose in IVR procedure. Evaluate ESD on IVR equipment as a reference to DRL settings, I would like to present the direction of improvement in the ESD rate test criteria for fluoroscopy dose. The experimental method is measured with 6cc ionization chamber under the 20cm PMMA Phantom. Radiation is subject to abdominal procedure. The average dose rate of the incident surface was 21.6 ± 11.4 mGy/min. The highest dose equipment was 58.5 mGy/min, and there was no equipment exceeding the domestic standard of 100 mGy/min. However, there were five units above 50 mGy/min. To reduce fluoroscopy dose, it is recommended to reduce pulse rate, The dose increases as the image receptor ages. It is recommended to modify the domestic inspection criteria to 50 mGy/min.