상세검색
최근 검색어 전체 삭제
다국어입력
즐겨찾기0
155189.jpg
KCI등재 학술저널

수술 중 급격히 발생한 친수성 다초점 인공수정체 혼탁 1예

A Case of Rapidly Occurring Intraoperative Clouding of Hydrophilic Multifocal Intraocular Lens

Purpose: To report a case of intraoperative clouding of hydrophilic multifocal intraocular lens (Lentis Mplus?? LS-313, Oculentis GmbH., Berlin, Germany) during cataract surgery. Case summary: A 69-year-old woman underwent intraocular lens exchange in her left eye due to clouding of an intraocular lens occurred intraoperatively. This phenomenon was caused by the rapid change in temperature of a very cold IOL being implanted into the eye. Conclusions: It is important to prevent a rapid temperature change during intraocular lens implantation. Particularly in the country that have winter season and fluctuating temperatures, careful caution will be necessary.

증례보고

고 찰

REFERENCES

로딩중