Iopamidol과 Ioversol 제제 조영제의 사용 후 보관 기간에 따른 ¹H-NMR 분석
¹H-NMR Analysis according to the Storage Period after the Use of Iopamidol and Ioversol Formulations Contrast Agents
- 한범희(Beom-Hee Han)
- 대한방사선과학회(구 대한방사선기술학회)
- 방사선기술과학
- 방사선기술과학 제44권 제3호
- 등재여부 : KCI등재
- 2021.06
- 239 - 245 (7 pages)
Since the side effects of contrast agents occur in many ways, hospitals must clearly identify the contrast agents and know how to manage them. Storage of contrast media It is common practice to store contrast media in a warm room to reduce viscosity and improve patient comfort. However, it is important to recognize that long-term storage at high temperatures can shorten the shelf life of a product. Therefore, in this study, the presence or absence of chemical changes according to the storage period after using the contrast medium was analyzed. As the analysis equipment, a 500 MHz Nuclear Magnetic Resonance Spectrometer of Bruker Avance (Germany) possessed by Korea Basic Science Institute (KOREA BASIC SCIENCE INSTITUTE; KBSI) was used. For the X-ray iodide contrast medium, Pamiray contrast medium of Iopamidol and Optiray contrast medium of Ioversol, samples remaining after 7 days, 20 days, 30 days, and 1 year were collected, By acquiring the spectrum from the 1H-NMR spectrometer of the contrast medium stimulated by It was compared and analyzed with the standard sample. In conclusion, depending on the storage period of the contrast medium, no peaks of physical and chemical changes were observed in both the Pamiray contrast medium of Iopamidol and the Optiray contrast medium of Ioversol after 7 days, 20 days, 30 days, and 1 year after use.
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론