몬테카를로 방법을 이용한 이동형 X선 투시검사 시 수술실 내 공간선량평가
Spatial Dose Distribution for C-arm Examination within Operation Room Using Monte Carlo Method
- 대한방사선과학회(구 대한방사선기술학회)
- 방사선기술과학
- 방사선기술과학 제44권 제3호
-
2021.06205 - 210 (6 pages)
-
DOI : 10.17946/JRST.2021.44.3.205
- 48
The purpose of this study was to analyze the spatial dose according to the distance by location of medical workers when using a mobile X-ray fluoroscopy device in the operating room through a simulation experiment. The MCNPX program was used for the simulation, and the location of medical workers was set around the operating table, and the spatial dose distribution according to the distance and changes in imaging conditions was evaluated. As a result, The highest score was 2.74×10-4 mGy, 2.72×10-4 mGy, and 1.18×10-4 mGy based on the 10 cm distance from the operating table. Spatial dose depending on the distance 100cm, A point 5.15×10-5 mGy is decreased 19% of 10cm, D point 5.12×10-5 mGy, 19 % of 10cm , and G pint, 1.73×10-5 mGy is reduced by 15% of 10cm. Based on this study, medical-related workers directly or indirectly participating in surgery carry potential risks of radiation exposure during surgery, but there are difficulties in radiation protection due to the nature of their work. Therefore, efforts to reduce exposure suitable for the operating room environment will be required.
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
(0)
(0)