Objective: In this study, the differences between two reconstruction methods were analyzed by comparing image uniformity and contrast according to Iteration and Subset, which were altered through the Iterative method and True X method, used in Siemens’ PET/CT studies. Methods: The Phantom images were obtained by exposure for two minutes per one bed. To determine the image uniformity, the Coefficient of variance was used. Also, in order to compare the contrast value, we measured and analyzed the ratio of the SUV mean of Phantom image’s hot spheres and the background. Results: Under the same reconstruction conditions (Iteration and Subset) of CV, the Iterative method was higher than the True X method. In the comparison of the SUV mean ratio of the background and hot sphere, the True X method had a closer rate than the Iterative method. Conclusion: The newly developed True X reconstruction method is better than the previously used Iterative method in terms of uniformity and contrast. However, the date for this study was only obtained using the Phantom device. In order to obtain more accurate and useful information from the experiment, further research should be conducted. Also, it is necessary to find the appropriate standards for Iteration and Subset for further experimentation.