Onco Flash에서 매개변수 변화에 따른 영상의 질 평가

Evaluation of Image Quality by Parameter Change in Onco Flash

  • 2
커버이미지 없음

Purpose: Many companies developed a lot of programs with continuous effort for program upgrading. With this acquire superior image quality for the purpose of quick examination and progress in spatial resolution. This study was to obtained clinical usefulness on a appropriate parameter of FWHM for speed and alpha value for superior image quality. Materials and Methods: Gamma camera by Siemens (e.cam) and spatial resolution phantom and four quadrant bar phantom used. Compared for FWHM by changed scan speed 15, 20, 25, 30, 35, 40 cm/min in scatter and non scatter in Onco Flash of spatial resolution phantom. Visual evaluation of count rate and bar phantom image for increased of alpha value of 10% in 0~100% . Results: FWHM by the scan speed was 9.37, 9.40, 9.28, 9.30, 9.31, 9.53 mm in the scatter. Count rate increased alpha value 10% increased. Visual evaluation was suitable to below 30%, Therefore spatial resolution improved on FWHM at the scan speed 25~35 cm/min applying for alpha value 30% in Onco Flash was average 9.3 mm less than FWHM of below 15 cm/min and above 40 cm/min. Conclusion: We found on appropriate parameter to progress of image quality. And there be a useful guideline for you that appropriate scan speed on vary in parameters of reduction on examination time and advancing image quality.