자동 분주기 Fixed Tip의 Carryover 방지를 위한 개선 방안
The Prevention of Carryover in the Automated System With Fixed Tips
- 대한핵의학기술학회
- 핵의학기술
- Vol.13 No.2
-
2009.10181 - 184 (4 pages)
- 2
Purpose: Automated system that immunoassay examination are used widely. However, sample to sample carryover can cause that the next patient sample is false positive. Materials and Methods: We test HBs Ag, HBs Ab, HBc Ab(IgG) with Automated pipetting system (Tecan). It was placed with very high concentrations followed by saline solution. During this experiment, The fixed tip of Automated system wash With 0.25 moL/L NaOH. The Measurement results of saline solution confirm the occurrence of carryover. Results: Results of saline solution with 0.25 moL/L NaOH cleaning process was measured 100% negative, And results of patient serum with 0.25 moL/L NaOH cleaning process is similar reported results. Conclusion: As Results, 0.25 moL/L NaOH cleaning process was avoid carryover in experiment, And we know results of the hepatitis test did not affected by this solution . we recommend 0.25 moL/L NaOH cleaning process as the Prevention of carryover in the automated system with fixed Tips.
서론
측정대상 및 방법
결과
고찰
결론
요약