장비에 따른 SUV의 차이와 이에 관한 고찰
Assessment and Comparison of SUVs of Three Different PET/CT Scanners
- 대한핵의학기술학회
- 핵의학기술
- Vol.15 No.1
-
2011.0434 - 38 (5 pages)
- 2
Purpose: The SUV is a widely used semi-quantitative index in PET for the estimation of radio-tracer accumulation in VOI. In this study, SUVs from three different PET/CT scanners were assessed, and differences between SUVs were evaluated. Materials and Methods: The PET/CT scanners which were assessed in this study were GEMINI, GEMINI TF 64 (Philips) and Biograph True Point True V 40 (Siemens). The NEMA PET phantom (Data Spectrum Corp., USA) was used to evaluate SUVs. The NEMA PET phantom has6.8 kg weight and three hot inserts. Two different activity distributions for the background and inserts were tested. The activity ratio were 3.7:3.7:7.4:11.1 MBq (1:1:2:3) and 1.85:7.4:9.25:11.1MBq (1:4:5:6) for each of background, insert 1, insert 2 and insert 3. Acquisition time was 2 minutes per bed position and NEMA PET phantom could be covered by two bed positions for all PET/CT scanners. The SUVs from each PET/CT scanner were compared with calculated true value. Results: For both activity ratios, all scanners showed similar results. The differences between each scanner were insignificant. Each scanner showed 91.2%, 85.9% and 87.2% of true SUV for GEMINI, GEMINI TF 64, Biograph True Point TrueV, respectively. Conclusion: For all scanners, SUVs were slightly lower than true value. However, the difference between scanners was insignificant. The SUVs from these scanners would be clinically meaningful if their consistent underestimation is kept in mind.
서론
재료 및 방법
결과
결론
요약