전신 뼈검사에서의 영상 평가 연구

Study on Image Quality Assessment in Whole Body Bone Scan

  • 5
커버이미지 없음

Purpose Whole body bone scan, which makes up a largest percentage of nuclear medicine tests, has high sensitivity and resolution about bone lesion like osteomyelitis, fracture and the early detection of primary cancer. However, any standard for valuation has not yet been created except minimum factor. Therefore, in this study, we will analysis the method which show a quantitative evaluation index in whole body bone scan. Materials and Methods This study is conducted among 30 call patients, who visited the hospital from April to September 2014 with no special point of view about bone lesion, using GE INFINIA equipment. Enumerated data is measured mainly with patient's whole body count and lumbar vertabrae, and the things which include CNR (Contrast to Noise ratio), SNR (Signal to Noise ratio) are calculated according to the mean value signal and standard deviation of each lumbar vertabrae. In addition, the numerical value with the abdominal thickness is compared to each value by the change of scan speed and tissue equivalent material throughout the phantom examination, and compared with 1hours deleyed value. Completely, on the scale of ten, 2 reading doctors and 5 skilled radiologists with 5-years experience analysis the correlation between visual analysis with blind test and quantitative calculation. Results The whole body count and interest region count of patients have no significant correlation with visual analysis value throughout the blind test(ρ<0.05). There is definite correlation among CNR and SNR. In phantom examination, Value of the change was caused by the thickness of the abdomen and the scan speed. And The poor value of the image in the subject as a delay test patient could be confirmed that the increase tendency. Conclusion Now, a standard for valuation has not been created in whole body bone scan except minimum factor. In this study, we can verify the significant correlation with blind test using CNR and SNR and also assure that the scan speed is a important factor to influence the imagine quality from the value. It is possible to be some limit depending on the physiology function and fluid intake of patient even if we progress the evaluation in same condition include same injection amount, same scan speed and so on. However, that we prove the significant evaluation index by presenting quantitative calculation objectively could be considered academic value.

서론

대상 및 방법

결과

결론 및 고찰

요약

REFERENCES

(0)

(0)

로딩중