범용 신틸레이터 감마카메라와 심근전용 반도체 감마카메라의 성능 비교 연구
Compared Performance of Semiconductor SPECT in Myocardial Perfusion SPECT: Phantom study
- 대한핵의학기술학회
- 핵의학기술
- Vol.20 No.2
-
2016.1049 - 53 (5 pages)
- 16
Purpose Recently, Cadmium-zinc-telluride (CZT) semiconductor myocardial SPECT (Single Photon Emission Computed Tomography) has been used myocardial scintigraphy. In this study, the performance of Semiconductor SPECT and conventional SPECT systems was compared by a comprehensive analysis of phantom SPECT images. Materials and Methods Methods: We evaluated the DSPECT CZT SEPCT (Spectrum-dynamic) and INFINA conventional (GE). Physical performance was compared on reconstructed SPECT images from a phantom. Results For count sensitivity on cardiac phantom images (counts · sec⁻¹ · MBq⁻¹), DSPECT had a sensitivity of conventional SPECT. This classification was similar to that of myocardial counts normalized to injected activities from phantom images (respective mean values, counts · sec⁻¹ · MBq⁻¹: 195.83 and 52.83). For central spatial resolution: DSPECT, 9.47mm; conventional SPECT, 16.90mm. For contrast-to-noise ratio on the phantom: DSPECT, 4.2; conventional SPECT, 3.6. Conclusion The performance of CZT cameras is dramatically higher than that of conventional SPECT. However, CZT cameras differ in that spatial resolution and contrast-to-noise ratio are better with conventional SPECT, whereas count sensitivity is markedly higher with the DSPECT.