Gamma Camera에 있어 측면 선란선의 영향에 대한 평가
The Evaluation of Lateral Scatter Ray of Gamma Camera
- 대한핵의학기술학회
- 핵의학기술
- Vol.22 No.1
-
2018.0546 - 50 (5 pages)
- 0
Purpose Generally, a collimator that installed in front of detector set a direction of gamma ray and remove a scatter ray. By the way, a lateral or oblique scatter ray is detected into crystal through collimator. At this study, we will evaluate a mount of count and spectrums of lateral scatter ray. Materials and Methods We used the SKY LITE (philips, netherlands) as a gamma camera, and ⁹⁹ᵐTc, 1.11 GBq point source as a phantom. we put this point source at backside 50 ㎝ of detector. After acquiring this for 1 min, we turned a detector next 10 degrees. Likely this, we acquired images at every 10 degrees from 0° to 360°, analyzed images and spectrums. In case of patient study, we choose a 3 phase bone scan patient who had a hand disease, because scatter rays from body would detect on crystal. After acquiring blood flow and blood pool images, we analyzed images and spectrums. Additional, we put a lead gown on patient’s hand, body. And then we compared and evaluated 3 type blood pool images (non lead gown, lead gown on a hand and on body). Results In case of phantom study, scatter ray counts at backside (270°-90°) are same with a background count. By the way, counts of scatter ray of oblique side (0°-50°, 220°-270°) are 100-600 cps, furthermore, counts at frontside are over 4 Mcps. In case of patient study, a counts of hand blood pool scan are 1510 cps. But counts of hand with lead gown on hands and on body are each 1554 cps, 1299 cps. Conclusion Therefore, even though there is a collimator in front of detector, lateral scatter rays detect on crystal and affect to images and spectrums. Especially, if there is a high activity source at outside of detector when we examine low activity organs like hands or foot, we have to shield and remove the source at outside for a good image.
서론
실험재료 및 방법
결과
결론 및 고찰
요약
REFERENCES
(0)
(0)