상세검색
최근 검색어 전체 삭제
다국어입력
즐겨찾기0
국가지식-학술정보

4MV 선형가속기에서의 비대칭 콜리메이터의 선량측정

Dosimetric Measurement for 4MV X-Ray Linear Accelerator with Asymmetric Collimator System

  • 0
커버이미지 없음

Dosimetric measurement of an asymmetric collimator system was performed, using water phantom system for 4MV X-ray linear accelerator. We have studied the system of dose calculation with those measured result. We compared the field size factor and the percent depth dose for asymmetric collimator to those factor for symmetric fields. The results show that we can use symmetric field data directly within 1% error, if we consider the off axis ratio(OAR).

(0)

(0)

로딩중