Average 기법에 의한 Visual-TEP의 검출과 DC-Drift 제거에 의한 TEP 신호개선에 관한 연구

Studies on the Detection of Visual-TEP with Average Method & the Improvement of TEP with DC-Drift Elinination

  • 0
커버이미지 없음

This paper presents average method to detect Visual-Transient Evoked Potential from the human scalp electric potential measured by the ElectroEncephaloGram. To confirm the validity of average method, the average-process is performed with both stimulated and no-stimulated potentials respectively, and both results are compared. The specific waveform, which is visual-transient evoked potential, is produced only in the case of stimulated potential. It was found that a dc-drift, due to instrumentation errors and other noises, can produce significant changes in the evoked-potential waveform. This can be removed with a high-pass filter (cut-off frequency=0.5Hz).