X선 진단영역 에너지 측정을 위한 Photo-Diode 선량계(PD-2000)의 개발
The development of photo-diode dosimeter(PD-2000) for the diagnostic X-ray Energy
- 대한방사선과학회
- Journal of radiological science and technology
- Vol.23 No.2
-
2000.0127 - 32 (6 pages)
- 0
It was produced radiation dosimeter used photo-diodes for which ionization by x-ray was applied and evaluated the value of utility in clinics as compared with ion-chamber. The result obtained were as follows : 1. Comparison of ion-chamber with photo-diode dosimeter's x-ray output by the change of x-ray tube voltage, and the ratio of ion-chamber to diode was $0.96{\sim}1.02$ which was not affected by x-ray beam quality. 2. The ratio of ion-chamber to diode was 0.96 by change of tube current and 0.97 by change of exposure time that is not affected by x-ray quantity. 3. The ratio of ion-chamber to diode was $0.97{\sim}1.04$ by thickness and $0.93{\sim}1.10$ by radiation field that is little affected by second ray quantity. 4. Reproducibility of photo-diode dosimeter was 0.011(CV) and it is a good result. 5. Photo-diode dosimeter was affected by the surface angle of detector over 30 degrees. Produced dosimeter was small, light, and meets good result compared with ionization chamber. It was expected come into wide use in clinic.