64채널 DROS 심자도 시스템을 위한 검출 회로의 잡음 특성
Noise Characteristics of Readout Electronics for 64-Channel DROS Magnetocardiography System
- 한국초전도학회
- Progress in Superconductivity
- Vol.7 No.1
-
2005.0146 - 51 (6 pages)
- 0
We have developed control electronics to operate flux-locked loop (FLL), and analog signal filters to process FLL outputs for 64-channel Double Relaxation Oscillation SQUID (DROS) magnetocardiography (MCG) system. Control electronics consisting of a preamplifier, an integrator, and a feedback, is compact and low-cost due to larger swing voltage and flux-to-voltage transfer coefficients of DROS than those of dc SQUIDs. Analog signal filter (ASF) serially chained with a high-pass filter having a cut-off frequency of 0.1 Hz, an amplifier having a gain of 100, a low-pass filter of 100 Hz, and a notch filter of 60 Hz makes FLL output suitable for MCG. The noise of a preamplifier in FLL control electronics is $7\;nV/{\surd}\;Hz$ at 1 Hz, $1.5\;nV/{\surd}\;Hz$ at 100 Hz that contributes $6\;fT/{\surd}\;Hz$ at 1 Hz, $1.3\;fT/{\surd}\;Hz$ at 100 Hz in readout electronics, and the noise of ASF electronics is $150\;{\mu}V/{\surd}\;Hz$ equivalent to $0.13\;fT/{\surd}\;Hz$ within the range of $1{\sim}100\;Hz$. When DROSs are connected to readout electronics inside a magnetically shielded room, the noise of 64-channel DROS system is $10\;fT/{\surd}\;Hz$ at 1 Hz, $5\;fT/{\surd}\;Hz$ at 100 Hz on the average, low enough to measure human MCG.