임프란트 표면 처리에 따른 공명주파수 변화에 대한 연구

COMPARISON OF RESONANCE FREQUENCY ANALYSIS BETWEEN VARIOUS SURFACE PROPERTIES

  • 0
커버이미지 없음

Purpose: The aim of this study is to compare the stability between Mg-incorporated implant, TiUnite and Machined implant. Materials and Methods: Premolars of 3 Mini pigs (24 months) were extracted. After 2 months later, total 27 fixtures of implants (9 of each design : Machined/ TiUnite/ Mg-incorporated) were inserted into the mandible of 3 mini-pig. Implant stability was estimated by RFA in installation to 2, 4 & 6 weeks. Statistical analysis of RFA values was performed with time and between groups using repeated measure ANOVA and turkey's multiple comparison test. Results: In analyzing the mean value for the observation periods, three types of implants yielded a slight decrease in RFA mean value after 2 week, followed by increase at 4-6 weeks. Mg incorporated oxidized implants demonstrated significantly higher RFA mean values at 6 weeks comparing other groups. The difference of RFA value with time and between groups was statistically significant. Conclusion: We concluded that Mg implants may reduce failure rates of clinical implants In the early period of bone healing and Mg implants may shorten the bone healing time from surgery to functional loading.