Kimplant에 관한 3차원 유한요소 분석적 연구

Three Dimensional Finite Element Analysis of Kimplant

  • 0
커버이미지 없음

In this study, the biomechanical characteristics of Kimplant were compared with that of Branemark implant by using three dimensional finite element analysis. Two finite element models were fabricated by inserting each implant into the bone model. The bone model was designed to have 18mm height, 13mm width and 15mm length. The size of each implant was planned to have 4mm width and 10mm length. A 200N force was applied on the center of abutment top in three directions - vertical, horizontal and oblique. After analyzing the stresses of fixture and surrounding bone, following results were obtained. 1. There was similar stress distribution between the two models. 2. The magnitude of maximum principal stress on the implant was similar between the two models but the location of maximum principal stress on the implant was different. 3. The magnitude and location of maximum principal stress on the surrounding bone was similar between the two models.

(0)

(0)

로딩중