A FEM Analysis for Acetabular Component with Negative Poisson's Ratio in Total Hip Arthroplasty
- 한국전산구조공학회
- Computational Structural Engineering
- Vol.8 No.4
-
1995.0117 - 23 (7 pages)
- 0
커버이미지 없음
Based on the present FEM study for negative Poisson's-ratio UHMWPE, the following conclusions seem expected. 1) Negative Poisson's-ratio UHMWPE transfers less stresses to the subchondral or peripheral iliac bone, compared to the conventional UHMWPE with Poission's-ratio. 2) Negative Poisson's-ratio cup reduces stresses in UHMWPE cup itself as well as metal backing, and subchondral bone. 3) The reduction in periacetabular mechanical stresses would significantly reduce the rate of fatigue failure and consequently reduce the incidence of aseptic loosening of the cup due to wear or bone resorption.